Precisión del SimpliFiber Pro

Test equipment vendors use the term "linearity" when describing the accuracy of a fiber power meter. Linearity is specified across the power range of the meter. In the case of the SimpliFiber Pro meter, this can be described as:

Linealidad de medida de alimentación
  • 850 nm: ±0,2 dB for power from 0 dBm to -45 dBm, ±0,25 dB for power < -45 dBm.
  • 1300 nm, 1310 nm, 1490 nm, 1550 nm, 1625 nm: ±0,1 dB for power from 0 dBm to -55 dBm. ±0,2 dB for power > 0 dBm and < -55 dBm.


It is also important to know the stability of the power source being used. In the case of the SimpliFiber Pro power sources, this can be described as:

Estabilidad de potencia de salida

  • 850/1300nm source: ±0,1 dB durante 8 horas
  • 1310/1550 nm source: ±0,25 dB durante 8 horas

    At 23°C ±2°C (73°F ±3.6°F), after 5 minutes of warm-up time.